Instructions for use Primus Infinity Empowered SW 4.5n 213
Remedying faults
symbol next to meas-
ured values
The apnea alarm for the
measured value concerned
has been disabled.
Some apnea alarms are
disabled automatically in
some ventilation modes.
For more details
see page 132.
For more details
see page 132.
, , or symbol
next to measured values
One or both alarm limits for
the measured value con-
cerned has/have been dis-
abled.
Set alarm limits,
see page 171.
Grayed out measured val-
ues
The specified accuracy
cannot be maintained.
Grayed out values The set value differs from
the delivered value.
Priority Message Cause Remedy