Technical data
256
Instructions for use Primus Infinity Empowered SW 4.5n
Cylinder pressure
1)
Measuring range 0 to 250 kPa x 100 (0 to 3600 psi)
Resolution of the displayed value 1 kPa x 100 (14 psi)
Accuracy ±4 % or ±6 kPa x 100 (±4 % or ±87 psi), the greater value ap-
plies
Pressure measurement (optional)
Respiratory pressure
Measuring range –20 to 80 cmH
2O
Resolution of the displayed value 5 cmH
2O
Accuracy at 60 L/min ±5 % of measured value or ±2 cmH
2O, the greater value ap-
plies
Flow measurement (hot-wire anemometry)
All values apply under ATPD conditions:
Flow
Measuring range –180 to 180 L per minute
Resolution of the measurement 0.1 L per minute
Accuracy at 60 L/min ±8 % of the measured value
tidal volume V
T
Measuring range 0 to 9999 mL
Resolution of the displayed value 1 mL
Accuracy ±8 % of the measured value or ±5 mL, the greater value ap-
plies
1) Valid for Silverline pressure reducers