EasyManua.ls Logo

Dräger Primus Infinity Empowered - Page 257

Dräger Primus Infinity Empowered
277 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use Primus Infinity Empowered SW 4.5n 257
Technical data
Delta V
T
Measuring range 0 to 9999 mL
Resolution of the displayed value 1 mL
Accuracy ±16 % or ±10 mL, the greater value applies
Volume VT
INSP
Measuring range 0 to 9999 mL
Resolution of the displayed value 1 mL
Accuracy ±8 % of measured value or ±5 mL, the greater value applies
Minute volume MV
Measuring range 0 to 99.9 L per minute
Resolution of the displayed value 0.1 L per minute
Accuracy ±8 % of the measured value or ±0.1 L per minute, the greater
value applies
Compliance C
PAT
Measuring range 0 to 250 mL/hPa (0 to 250 mL/cmH2O)
Resolution of the displayed value 0.1 mL/hPa (0.1 mL/cmH
2O)
Accuracy ±15 % of the measured value or ±0.5 mL/hPa
(±0.5 mL/cmH
2O), the greater value applies
MV
LEAK
Measuring range 0 to 9.99 L per minute
Resolution of the displayed value 0.01 L per minute
Accuracy ±15 % of (MV
EXP + MVLEAK) or ±0.1 L per minute, the
greater value applies
MV
MAND
Measuring range 0 to 99.9 L per minute
Resolution of the displayed value 0.1 L per minute
Accuracy ±8 % of measured value or ±0.1 L per minute, the greater
value applies