2. Replacement and adjustment of parts
27810
4/4
27810
Replenishment package unit
2. Check the output amount.
For the operation procedures, see Operator's Manual - Additional Operations & Setup Manual -.
IMPORTANT
Make sure that the difference between the measured value and the output amount entered in the system is
within ±3%. If the difference is not within ±3%, check for damage or clogging of the pipes. If the pipes are fine,
there must be air remaining in the pumps. Exhaust the air, and carefully measure the output amount again.
When the output amount for each replenishment package is measured, the remaining amount of each solution
varies.
3. Replace the replenisher pipe and the water supply pipe to their original position.
4. Mark the new pump output amounts on the label.
(The label is affixed to the inside of processor door 1.)
5. Reattach the sub-tank top cover and processor cover 1.
IMPORTANT
If you do not clean the air exhaust tool after use, the syringe packing may be damaged.
Be sure to clean the tool before storing.