4. Troubleshooting
40900
10/13
40900
Attention message: F replenishment
No. 00913[F]
Mixing Replenisher cannot be started.
Attention message release
YES
Condition
NOTE
• If any one of the (lower) replenishment solution level sensors of P1R/P2RA/P2RB is on after clicking Mixing Replenisher and a
specified time has passed,
☞No. 05923[F] occurs.
Check Point
Diagnosis
NOTE
• To access the wiring diagrams, click Wiring diagrams in Bookmarks.
Condition
If the (lower) replenishment solution level sensors of P1R/P2RA/P2RB are on after the Mixing Replenisher is clicked
Suffix number Condition
00001 P1R
00002 P2RA
00004 P2RB
For details of operation specification, see ☞ 57010.
Countermeasure
1. Drain replenishment solution completely from the replenishment tanks of P1R, P2RA and P2RB. ☞27550
2. Perform Mixing Replenisher with each replenishment tank empty.☞33003
Check the status of # # # (lower) replenishment solution level sensors of each processing solution with
Input Check.
☞35220
Failed part Manual No.
P1R (Lower) Replenishment Solution Level Sensor
P2RA (Lower) Replenishment Solution Level Sensor
P2RB (Lower) Replenishment Solution Level Sensor
☞63280
F replenishment I/O PCB ☞66800
Processor control PCB
☞66200