4. Troubleshooting
45500
7/14
45500
Error message: Processor 1
No. 05509
The circulation amount has decreased.
Alarm release
NO
Error message release
YES
NOTE
This error message will be shown only when the system is equipped with the optional digital flowmeter.
Condition
NOTE
This error will not be displayed when the digital flowmeter is set to OFF via Option Registration.
Check Point
Diagnosis
NOTE
To access the wiring diagrams, click Wiring diagrams in Bookmarks.
Suffix number Condition
00000 The circulation amount of the P1 processing solution measured by the digital flowmeter falls below 3L/min.
Although the system is not equipped with a digital flowmeter, the digital flowmeter is set to ON via Option
Registration.
1 Check that the filter cartridge is not soiled.
2 Check if the circulation pump and the hose connected to the pump are clogged.
Failed part Manual No.
Digital flowmeter 63290
Processor control PCB
66200
Processor relay PCB 66220