45500
Error message: Processor 1
No. 05509
The circulation amount has decreased.
Alarm release
NO
Error message release
YES
NOTE
• This error message will be shown only when the system is equipped with the optional digital flowmeter.
Condition
NOTE
• This error will not be displayed when the digital flowmeter is set to OFF via Option Registration.
Check Point
Diagnosis
NOTE
• To access the wiring diagrams, click Wiring diagrams in Bookmarks.
Suffix number Condition
00000 The circulation amount of the P1 processing solution measured by the digital flowmeter falls below 3L/min.
Although the system is not equipped with a digital flowmeter, the digital flowmeter is set to ON via Option
Registration.
1 Check that the filter cartridge is not soiled.
2 Check if the circulation pump and the hose connected to the pump are clogged.
Failed part Manual No.
Digital flowmeter ☞63290
Processor control PCB
☞66200
Processor relay PCB ☞66220