ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-37
PRELIMINARY
7-4-3-3 RFI FE Cache Test (cont’d)
When the RFI FE Cache Test sequence is complete
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-15 "View Test Log - RFI
Cache Test" on page 7-37.
8.) If necessary, use the scroll bars and/or arrows to scroll through the list of Test results.
NOTE: For this test, there is one report (FE Cache Test report) which is listed in the Reports window
on the right. By default, this report is displayed in the Reports window below.
9) When done, trackball to the Exit button and press Set to close the View Report window.
10) After reviewing the test results, proceed as follows:
If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
If no problems were identified but the problem persists, contact your local GE HEALTHCARE
field engineer for assistance.
If errors were identified but the results were marginal, perform calibration, as described in
“DC Offset Calibration” on page 7-27, and then repeat the test.
If “Fail” appears in one of the Result columns, the test has failed, even if “Pass” appears in other
columns. Repeat the RFI FE Cache Test.
If errors were identified and the results were not marginal, verify that all system cables and
boards are connected (not misplaced) and functioning properly, as described in Chapter 5 -
Components and Function (Theory).
Figure 7-15 View Test Log - RFI Cache Test