ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
7-40 Section 7-4 - Performing Front End (FE) Diagnostics
PRELIMINARY
7-4-3-4 RFI Analog Circuits Test (cont’d)
.
8.) If necessary, use the scroll bars and/or arrows to scroll through the list of Test results.
NOTE: For this test, there is one report (Analog Circuits Test report) which is listed in Reports on the
right. By default, this report is displayed in the Reports window below (Figure 7-17 "View Test
Log - RFI Analog Circuits Test" on page 7-40). Use the Report window scroll bars and/or arrows
to scroll through complete report if necessary.
9) When done, trackball to the Exit button and press Set to close the View Report window.
10.)After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If no problems were identified but the problem persists, contact your local GE HEALTHCARE
field engineer for assistance.
• If errors were identified but the results were marginal, perform calibration, as described in
“DC Offset Calibration” on page 7-27, and then repeat the test.
• If errors were identified and the results were not marginal, verify that all system cables and
boards are connected (not misplaced) and functioning properly, as described in Chapter 5 -
Components and Function (Theory).
Figure 7-17 View Test Log - RFI Analog Circuits Test