EasyManua.ls Logo

GE Vivid S5 N - Page 432

GE Vivid S5 N
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
7-48 Section 7-4 - Performing Front End (FE) Diagnostics
PRELIMINARY
7-4-4 TR 32 Diagnostic Tests (cont’d)
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
The TR32 BF Digital Test commences. While the test proceeds, the name of the test currently in
progress is displayed in the Execute field and the current Loop # and specific test are shown below.
The messages Init Done and Please Wait are displayed beside the progress bar in the Status area:
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
Green: All recommended tests for these components have been completed and no problems
were identified.
Red: Problems were identified in these components during the performance of the test.
Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the TR32 BF Digital Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-23 "View Test Log - TR32
BF Digital Test (MLA0)" on page 7-48.
8.) If necessary, use the scroll bars and/or arrows to scroll through the list of Test results.
Figure 7-23 View Test Log - TR32 BF Digital Test (MLA0)

Table of Contents