GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
7-104 Section 7-5 - Monitoring Tests
PRELIMINARY
7-5-1              Accessing the Monitoring Diagnostic Test Options (cont’d)
• Green: All recommended tests for these components have been completed and no problems 
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all 
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Fans Speed Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details 
of which tests passed or failed, as shown in the example in Figure 7-70 "View Test Log - Fans 
Speed Test" on page 7-104.
NOTE: For this test, there is one report (Fans Status) which is listed in the Reports window on the right. 
By default, this report is displayed in the Reports window below.
7.) Check that the test passed and that result values for each fan are within the pre-defined Max and 
Min limits, as shown in Figure 7-70 "View Test Log - Fans Speed Test" on page 7-104.
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when 
prompted). 
• If no problems were identified but the problem persists, contact your local GE HEALTHCARE 
field engineer for assistance.
Figure 7-70 View Test Log - Fans Speed Test