ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
7-118 Section 7-6 - Performing Back End Diagnostics on the System
PRELIMINARY
7-6-1 Accessing the Back End Diagnostic Test Options (cont’d)
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Full PC Doctor Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests and sub-tests passed or failed, as shown in the example in Figure 7-78 "View
Test Log - Full Back End Test w/o Hard Disk Scan" on page 7-116.
NOTE: For the Full PC Doctor Test, many tests are performed in the test sequence; these are listed in
the Tests window as shown in Figure 7-80 "View Test Log - Full PC Doctor Test" on page 7-118.
For each currently-selected test (for example, CPU Full Test) the corresponding report is listed
in Reports on the right.
For details of each report, refer to the explanations provided in 7-6-1-1-3 "CPU Tests" on page
7-119:
7.) Use the scroll bars and/or arrows to scroll through the list of Test results and review the reports, as
required.
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
If no problems were identified but the problem persists, contact your local GE HEALTHCARE
field engineer for assistance.
If errors were identified and the results were not marginal, verify that all system cables and
boards are connected (not misplaced) and functioning properly, as described in Chapter 5 -
Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Contact your local
GE HEALTHCARE field engineer for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
Figure 7-80 View Test Log - Full PC Doctor Test