GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-9
PRELIMINARY
8-2-3 Left Side Cover Replacement Procedure (cont’d)
7.) Make sure the air filter is seated in the correct position then fasten the other end of the air filter onto
the cover, by closing the securing clip - refer to Figure 8-6 on page 8-7 and Figure 8-9 below.
8.) Return the left side cover to the system, as described in the “Left Side Cover Installation Procedure”
on page 8-7.
9.) Proceed to perform the following functionality tests:
- “Grounding Continuity” on page 10-23
- “Chassis Current Leakage Test” on page 10-24
- “Temperatures Test” on page 7-15
- “Fans Speed Test” on page 7-15
- “Fans Speed Setting Test” on page 7-15
- “Air Filter Cleaning” on page 10-11
Figure 8-9 Air Filter Fastened in Position on the Left Cover