GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-188 Section 8-6 - Lower Section Components Replacement Procedures
PRELIMINARY
Note: When reconnecting the ground leads, there are two spring washers that should be installed in
the order shown in Figure 8-237. In addition, place the ground lead connector with the flat
surface facing towards the Cart to prevent damage to the cable holder and the cable.
10.)Return the bottom rear cover to the system, as described in “Bottom Rear Cover Installation
Procedure” on page 8-23.
11.)Turn ON power to the system.
12.)Proceed to perform the following functionality tests:
- “Grounding Continuity” on page 10-23
- “Chassis Current Leakage Test” on page 10-24
- “Calibration” on page 7-12
- “Full System Test” on page 7-11
Figure 8-237 Ground Lead Showing Spring Washers and Cable Holder