Alarms, Action and Reminder Notications 1313 Alarms, Action and Reminder Notications
Diagnostics
Warning: DO NOT continue using an activated Pod that fails to beep
during a diagnostic test. e Pod should be changed immediately. If the
Omnipod 5 App fails to beep during a diagnostic test, contact Customer Care
immediately. Continuing to use the Omnipod 5 System in these situations
could put your health and safety at risk.
Check alarms
Before you begin, switch to Manual Mode if currently using Automated Mode.
See “22.2. Switching from Automated Mode to Manual Mode” on page 276.
Note: Your Controller sound settings must be ON to hear a tone on the
Omnipod 5 App.
To verify that your Omnipod 5 App and Pod's alarms and vibration functions are
working properly, test them as follows:
1. Navigate to: Menu button ( ) > Settings > General > Check Alarms.
2. If you have an active Pod, tap PAUSE INSULIN and tap YES.
3. Tap CHECK ALARMS to initiate the alarm check.
4. Listen and feel: e Controller beeps and vibrates three times. If you are
wearing a Pod, the Pod beeps several times and sounds the alarm tone for
several seconds.
5. If the Pod did not beep, tap NO. en either tap CHECK AGAIN to retry testing
the alarms, or tap DONE and change your Pod.
6. If the Omnipod 5 System alarms did not sound, tap NO. en either tap
CHECK AGAIN to retry testing the alarms or contact Customer Care.
7. If the beeps and vibrations worked properly, tap YES. If you paused insulin to
check the alarms, tap YES to start insulin delivery.