EasyManua.ls Logo

OmniPod 5 - Bolus Settings; Impacts to Suggested Bolus Calculations

OmniPod 5
366 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Delivering a Bolus with the SmartBolus Calculator 17
222
17 Delivering a Bolus with the SmartBolus Calculator
17.8. Bolus Settings
Warning: DO NOT start to use your system or change your settings without
adequate training and guidance from your healthcare provider. Initiating and
adjusting settings incorrectly can result in over-delivery or under-delivery of
insulin, which could lead to hypoglycaemia or hyperglycaemia. Settings that
impact insulin delivery mainly include: Pod Shutdown, basal rate(s), Max
Basal Rate, Max Bolus, Correction Factor(s), Insulin to Carb (IC) Ratio(s),
Minimum Glucose for Calculations, Target Glucose, Correct Above and
Duration of Insulin Action.
ese settings allow you to change your Maximum Bolus, extended bolus and
SmartBolus Calculator settings.
Impacts on Suggested Bolus Calculations
e following table describes what each SmartBolus Calculator setting does, how
you can adjust them and how they are used to calculate a suggested bolus.
Omnipod 5 Setting and
Range
How to Enter the
Setting
Impacts on Suggested
Bolus Calculations
Carbs (grams)
0.1 – 225 g
(0.1 g increments)
Enter in SmartBolus
Calculator
Increase in carb amount
value increases amount
of suggested bolus dose.
Decrease in carb amount
value decreases amount
of suggested bolus dose.
Sensor Glucose
Value (mmol/L)
2.2 – 22.2 mmol/L
(0.1 mmol/L
increments)
Select USE SENSOR
within SmartBolus
Calculator
(Value comes from your
connected Sensor)
Increase in
sensor glucose
value increases amount
of suggested bolus dose.
Decrease in
sensor glucose
value decreases amount
of suggested bolus dose.
Blood Glucose
Reading (mmol/L)
1.1 - 33.3 mmol/L
(0.1 mmol/L
increments)
Enter in
SmartBolus Calculator
(Value comes from your
blood glucose meter)
Increase in BG Reading
increases amount of
suggested bolus dose.
Decrease in BG Reading
decreases amount of
suggested bolus dose.

Table of Contents