223
Delivering a Bolus with the SmartBolus Calculator 1717 Delivering a Bolus with the SmartBolus Calculator
Omnipod 5 Setting and
Range
How to Enter the
Setting
Impacts on Suggested
Bolus Calculations
Maximum Bolus
0.05 – 30 U
(0.05 U increments)
Enter in Omnipod 5
App Settings or during
First-Time Set-up
Limits amount of single
bolus dose.
Extended Bolus
(Manual Mode only)
ON/OFF
Enter in Omnipod 5
App Settings or during
First-Time Set-up
Allows for bolus delivery
over a user-selected
period of time.
Target Glucose &
Correct Above
Target Glucose:
6.1–8.3mmol/L
Correct Above:
6.1–11.1 mmol/L
(0.55 mmol/L
increments, up to
8segments/day)
Enter in Omnipod 5
App Settings or during
First-Time Set-up
Increase in setting
value decreases amount
of suggested bolus dose.
Decrease in setting
value increases amount
of suggested bolus dose.
Minimum Glucose for
Calculations
2.8 – 3.9 mmol/L
(0.1 mmol/L
increments)
Enter in Omnipod 5
App Settings
Disables SmartBolus
Calculator when glucose
is at or below setting
value.
Insulin to Carb Ratio
1 – 150 g
(0.1 g increments, up to
8segments/day)
Enter in Omnipod 5
App Settings or during
First-Time Set-up
Increase in setting
value decreases amount
of suggested bolus dose.
Decrease in setting
value increases amount
of suggested bolus dose.
Correction Factor
0.1 – 22.2 mmol/L
(0.1 mmol/L increments,
up to 8 segments/day)
Enter in Omnipod 5
App Settings or during
First-Time Set-up
Increase in setting
value decreases amount
of suggested bolus dose.
Decrease in setting
value increases amount
of suggested bolus dose.
Reverse Correction
ON/OFF
Enter in Omnipod 5
App Settings
If “On, suggested bolus
is decreased when
glucose is below Target
Glucose
value.