Browsing Your History and Records 11
132
11 Browsing Your History and Records
11.5. Calculations for History Summaries
is section lists the calculations for the summary data shown on the History
screens.
Glucose Summaries
e calculations used for the glucose summaries include sensor glucose values
(including HIGH and LOW values) and exclude manually-entered readings.
Item Calculation
Row 1
Average
Sensor
=
Sum of all sensor glucose values
Total number of sensor glucose values
Note: HIGH sensor glucose values are included as 401 and LOW
sensor glucose values as 39.
Sensor
in
Range
=
Qty of sensor glucose values within Glucose Goal Range
x 100
Total number of sensor glucose values
Sensor
Above
=
Qty of sensor glucose values above Glucose Goal Range upper limit
x 100
Total number of sensor glucose values
Sensor
Below
=
Qty of sensor glucose values below Glucose Goal Range lower limit
x 100
Total number of sensor glucose values
Insulin Delivery Summaries
Bolus insulin calculations include:
SmartBolus Calculator boluses
manual boluses
any partial bolus amounts delivered from immediate or extended boluses
that were cancelled that the Pod can conrm
Basal insulin calculations include:
basal adjustments from SmartAdjust technology (Automated Mode)
Basal Programmes (Manual Mode)
any adjustments for temp basals, pause periods or times without an active
Pod