Alarms, Action and Reminder Notications 13
158
13 Alarms, Action and Reminder Notications
Pod Shutdown
Caution: ALWAYS activate a new Pod in a timely manner. Waiting too long
between Pod changes could result in under-delivery of insulin, which can lead
to hyperglycaemia. If another Pod is not available, use a dierent insulin delivery
method.
Screen Alert Description
Omnipod 5 App: Why it occurs: e Pod has stopped
delivering insulin because you have set a Pod
Shutdown time and you did not respond to
the Pod Shutdown advisory alarm.
Pod sound: Continuous tone
Controller sound and vibration:
• 30-second tone
• 30-second vibration
• Vibration and tone repeat every
15 minutes until acknowledged
What to do:
1. Tap OK, DEACTIVATE POD NOW.
2. Change your Pod.
3. Check your blood glucose.
Note: To change the Pod Shutdown setting,
see “Pod Shutdown” on page 121.
Lock Screen: