EasyManua.ls Logo

OmniPod 5 - Page 223

OmniPod 5
366 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
215
Delivering a Bolus with the SmartBolus Calculator 1717 Delivering a Bolus with the SmartBolus Calculator
Additional information
You may choose to accept or change the nal recommendation before the bolus is
delivered.
Aer opening the SmartBolus Calculator, bolus delivery must be initiated within
5 minutes or values will need to be refreshed. If more than 5 minutes pass, you will
see a message that values have expired. Tap CONTINUE to refresh the SmartBolus
Calculator, then enter or use your current values.
To change your personal settings used by the SmartBolus Calculator, see
SmartBolus Calculator Settings on page 224.
Note: To use the SmartBolus Calculator, the Omnipod 5 App and the Pod must
be communicating. If there is no Omnipod 5 App-to-Pod communication, you
will be prompted to re-establish a Pod connection. To nd out what to do when
your Omnipod 5 App and Pod have a communication issue, see “Frequently
Asked Questions and Troubleshooting on page 311.
Caution: DO NOT navigate away from the Omnipod 5 App while you are
in the process of making changes to your insulin delivery settings. If you leave
the App before you are able to save the setting change and before the App is
able to put the setting change into eect, the system will continue to use your
last saved settings. As a result, you may continue with therapy settings that
you did not intend. If you are unsure about whether your changes were saved,
review your settings.
Tip: If you already know the number of units for the bolus you want to deliver, tap
Total Bolus. Enter the bolus amount and tap Done. en go to “17.6. Delivering
an Immediate Bolus on page 219.