EasyManua.ls Logo

OmniPod 5 - Page 348

OmniPod 5
366 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
340
Appendix
Caution: DO NOT use portable radio frequency (RF) communications
equipment (including peripherals such as antenna cables and external
antennas) closer than 30 cm (12 inches) to any part of the Omnipod 5 System,
as it may impact the communication between your Controller and your Pod.
Electromagnetic Emissions
is equipment is intended for use in the electromagnetic environment
specied below. e user of this equipment should ensure that is used in such an
environment.
Emissions Compliance
according to
Electromagnetic environment
RF Emissions
(CISPR11)
Group 1 e Pod, Controller, and Transmitter emit
low-level electromagnetic energy (RF) in order
to communicate. Although unlikely, nearby
electronic equipment may be aected.
CISPR B
Emissions
Classication
Class B e System is suitable for use in all establishments,
including domestic establishments.
Harmonic
Emissions
(IEC 61000-3-2)
Class A
Voltage
Fluctuations/
Flicker Emissions
(IEC 61000-3-3)
P
st
≤ 1.0
P
lt
≤ 0.65
d
c
≤ 3%
d
max
≤ 4%
d
(t)
≥ 200 ms
during a voltage
change should be
≤ 3%
Electromagnetic Immunity
e System is intended for use in the electromagnetic environment specied
below. You should observe these requirements in the use of the System.
Immunity
against
IEC 60601-1-2
test level
Compliance
level (of this
device)
Electromagnetic environment
ElectroStatic
Discharge, ESD
(IEC 61000-4-2)
contact
discharge:
± 8 kV
air discharge:
± 15 kV
± 8 kV
± 15 kV
If oors are covered with synthetic material,
try to avoid electrostatic discharges.

Table of Contents