52
3 Omnipod 5 System Overview
Identifying System modes
e mode indicator shows the Omnipod 5 System's current operating mode.
Graphic Description
Displays when there is no Pod
communication or no active Pod.
Displays when the Omnipod 5 System is in
Automated Mode and the Pod is providing
automated insulin delivery.
Displays when the Omnipod 5 System is
in Automated Mode: Limited state. e
most common reason is that the Pod is not
receiving sensor glucose values. In response,
the system is delivering basal insulin based
on a calculation of user-entered settings
and past insulin delivery. Check your
Sensor to make sure it is functioning. e
position of the Pod and Sensor may also
be contributing to the loss of connectivity
between the devices.
Displays when the Omnipod 5 System is
in Manual Mode and delivering the active
Basal Programme.