Performing internal logging
ZW-7000/5000
User's Manual
107
3
Basic Operation
Item Setting item Range Description
Data logging
conditions
Data interval 1 to 1000 Set the data storage interval.
If “1” is set , all measured data is stored, and “2” is set, one
measurement
value is stored every two measurement.
If 0 is set, data is saved only when the hold value is applied.
Number of
logging da
ta
1 to 2000000 Set the number of data to store for the data that is assigned
to each OUT.
OUT1 OFF/ON Set whether to store the measurement results for OUT1.
OUT2 OFF/ON Set whether to store the measurement results for OUT2.
OUT3 OFF/ON Set whether to store the measurement results for OUT3.
OUT4 OFF/ON Set whether to store the measurement results for OUT4.
2
Click the [Logging start] button to start
internal logging.
A confirmation message is displayed. Click
[Y
es]
to start internal logging.
When the label insert mode is OFF, the internal
logging
data is automatically cleared before
internal logging is started.
Once internal logging starts, another internal
l
ogg
ing
cannot be started until it ends.
• Starting internal logging can also be executed by No-protocol communications or parallel I/O.
• For assigning of OUT1 to OUT4, refer to th
e following:
“2-1 Parallel I/O connection” described in Displacement Sensor ZW-7000 series Conf
ocal Fiber Type
Displacement Sensor User’s Manual for Communications Settings (Z
363).
3
After internal logging ends, click the [Save to
file (Sensor → PC)] to output the data to a
file.
A
CSV format file in the following format is
o
utput
. Unit of each output values are the
following:
Measure value: nm
Amount of emitted light: 0.01% (0.01 to
1
00.00%)
Amoun
t of received light: Luminance level (0 to
4
095)
Note