Displaying measured values in graphs
ZW-7000/5000
User's Manual
97
3
Basic Operation
Specifying the sampling start and end conditions
You can specify the conditions for starting and ending sampling.
1
Check the Trigger start conditions/Trigger end conditions checkbox.
2
Select the trigger condition.
Item Setting item Range Description
Trend setting Trigger start
conditions
I/O Specify parallel I/O
Specify parallel I/O: written the belows, as the trigger
conditi
on.
Parallel I/O are specified as the trigger condition the
follo
wing:
TIMING, ZERO, BUSY, ENABLE, HIGH, PASS, LOW,
TASKSTAT, LOGGING, LOGSTAT, LOGERR, STABILIT as
the
tr
igger condition.
Data slope Sets the change in the measurement values for OUT1 -
OUT4 as the star
t condition. Sets the trigger level. The
time the measurement value rises above this value or falls
below it is set as the start condition.
Data window (In) Sets the change in the measured values for Out1 to Out4
as t
he sta
rt condition. When the measured value is inside
the range, the start trigger is issued.
Data window (Out) Sets the change in the measured values for Out1 to Out4
as the star
t condition. When the measured value is outside
the range, the start trigger is issued.
Trigger end
condition
s
I/O Specify parallel I/O (TIMING, ZERO, BUSY, ENABLE,
HIGH, PASS, LOW, SYNC/TRIG, SYNCFLG,
TASK_ENABLE, LOGGING, LOGSTAT, LOGERR,
STABILITY RESET,BUFFER_ERROR) as the trigger
condition.
Data slope Sets the change in the measurement values for OUT1 -
OUT4 as
the start condition. Sets the trigger level. The
time the measurement value rises above this value or falls
below it is set as the start condition.
Data window (In) Sets the change in the measured values for Out1 to Out4
as t
he sta
rt condition. When the measured value is in the
range, the end trigger is issued.
Data window (Out) Sets the change in the measured values for Out1 to Out4
as the star
t condition. When the measured value is outside
the range, the end trigger is issued.
Number of data The sampling data is counted from when the start
co
nditi
on is met and when the specified number of data
points have been sampled, the end trigger is issued.