Setting the Output Conditions
ZW-7000/5000
User's Manual
225
Sensor controller operations
7
Setting Hold
Setting the Hold Mode
Setting the Hold Mode p.144
As an example, here is an explanation of the procedure fo
r selecting the “PEAK” hold mode setting.
Operating procedure
Steps Key operation Display Description
1 Press and hold the key for two seconds to enter the FUNC
mode.
2 Press the / key to select MEAS, then press the key.
3 Press the / key to select TASK1, then press the key.
4 Press the / keys to select “OUTPUT” and press the
key.
5 Press the / keys to select “HOLD” and press the
key.
6 Press the / keys to select “TYPE” and press the
key.
7 Set the hold mode.
PEAK: Peak
BOTTOM: Bottom
P-P: PEAK TO PEAK
AUTOPK: AUTO PEAK
AUTOBT: AUTO BOTTOM
AUTOPP: AUTO PEAK TO PEAK
AVE: AVERAGE
SAMPLE: Sampling
OFF: Through
The current setting value is displayed on the sub-display.
Press / key to enter editing mode and the sub-display blinks.
In this example, select “PEAK” and press the
key.
8 Press the key for two seconds to enter the RUN mode.