Setting the System
244
ZW-7000/5000
User's Manual
Calibrating Sensor Head
Calibrating Sensor Head p.66
Operating procedure
When Sensor Head calibration fails
If the Sensor Head is calibrated in an inappropriate environment, an error is displayed on the main display. In such a
case, return to the previous screen using key. Then refer to the p.66, recheck your environment, and try again.
If an error results even if Sensor Head calibration is done under the correct conditions
, it is possible that some
contaminants may have adhered to the Sensor Head part, or the fiber connector of the Sensor Controller. Clean the
fiber cable or fiber connector referring to p.59.
Steps Key operation Display Description
1 Press and hold the key for two seconds to enter the FUNC
mode.
(For details on the functions of the Mode switching key, see p.197
.)
2 Press the / keys to select “SYSTEM” and press the key.
3 Press the / keys to select “H.CALIB”, then press the key.
4 “OK/CAN” is displayed on the sub-display.
Press the key to execute calibration or the
key to cancel.
5 Press the key for two seconds to enter the RUN mode.