Basic Operations for Displaying Measurement Results
ZW-7000/5000
User's Manual
25
1
Basic configuration
1-3 Basic Operations for Displaying Measurement Results
This section describes a series of operations, from setting up the Sensor Head and Sensor Controller
and connecting to the PC tool (Sysmac Studio), to calibrating the Sensor Head, through to monitoring the
measurement results. For details, refer to the description of their respective chapters.
Installation of Sensor Head
Installation procedure
2 Installation and Connections p.39
Attaching the strap to the Calibration ROM
1
Place the Sensor Head with an appropriate distance from
the target to measure, fixing it by tightening four M3 screw
inserted into their respective installation holes.
Tightening torque: 0.54 N • m
9-1 Specifications and External Dimensions p.260
1
Attach the supplied strap to the calibration ROM.
M3 screw×4
Calibration ROM
Strap