Performing internal logging
108
ZW-7000/5000 
User's Manual
Even when the label insert mode is ON, the label data is not output.
Performing Internal Logging with Parallel I/O
The logging start and end timings can be specified using LOGGING input signals from parallel I/O. When doing 
so, follow the procedure below to set the data logging conditions.
 Multi view Explore : [System] (double-click) 
→ Edit pane : [Sensor Settings] icon ( ) 
→ Sensor settings window : [Internal logging setting]
1
Set the logging conditions for internal logging.
There are three settings: Number of logging, number of log
ging decimation, and overwrite mode.
LOGERR signal is turned ON at the following cases:
• Overwrite mode is OFF
• 2,000,000 items have already logged in the internal memory at the begging of logging.
When continue to logging regardless number of logging while LOGGING input signal is turned ON, set 
as
 bel
ow:
• Number of logging data: 0
• Overwrite mode: ON
LoggingMode  Internal
SamplePeriod 1
SamplingCycleTime 0.2
Index OUT1 OUT2 OUT3 OUT4
0 -1938 2823 9437 -2147483648
1 -1939 2764 9401 -2147483648
2 -1940 2773 9355 -2147483648
3 -1940 2807 9523 -2147483648
4 -1941 2794 9442 -2147483648
5 -1941 2817 9497 -2147483648
Item Setting item Setting value Description
Internal logging 
setting
Number of 
logging data
0 to 2000000
(default value: 100000)
Set the number of data to save for each internal logging 
output data.
Internal logging 
settings
0 to 1000
(d
efault value: 1)
Set the interval of saving internal logging data.
If 1 is set, all data is saved. If 2 is set, one measured data is 
save
d every two measurements. 
If 0 is set, data is saved only when the hold value is applied.
Overwrite mode OFF (default value)/ON Set whether to make the internal logging mode run in 
o
ve
rwrite mode.
Sets the operation to be performed when the logging count 
excee
ds the set number of logging data. 
OFF: Internal logging is stopped. 
ON: The oldest logging data is overwritten.
Label insert 
mod
e
OF
F (default value)/ON Set the operation to be performed when logging input is 
turned on.
OFF: No operation.
ON: When logging input is turned on, inserts label into log-
ging data.