Setting the Output Conditions
130
ZW-7000/5000
User's Manual
4-3 Setting the Output Conditions
Setting Scaling
This setting is used when you want to correct any errors that are generated due to the installation status of the
Sensor Head, and display the corrected value on the main display as a measured value.
There are three types of setting: “auto scaling” (“1-point scaling” and “2-point scaling”) that automatically sets
th
e correction
value of a placed sensing object, “fixed scaling” that manually sets the correction value and
“thickness scaling” that automatically sets the correction value of a placed sensing object (transparent object).
The "Zero reset" p.92 settings return to the default settings when scaling is set. If scaling is set, perform a zero reset
as necessary.
Performing Manual Scaling
Set the scaling by entering the correction value.
Item Setting item Setting value Description
Scaling
(Manual scaling)
Scaling OFF (default value)/ON Set scaling ON/OFF.
Span -10.0000 to 10.0000
(default value: 1.0000)
Set the inclination of the sensor character
istics as a
coefficient.
Offset -999.999999 to 999.999999
(default value: 0)
A fixed value is added to/subtracted from the measured
valu
e.
Important
-10.0
10.0
Measurement value
(mm)
Workpiece
displacement
Measurement value
(mm)
Workpiece
displacement