Setting I/O
ZW-7000/5000
User's Manual
159
4
Settings for Function
4-5 Setting I/O
Settings for Processing When Measrement Is Not Possible
Setting operation when measurement is not possible
Set the output method when the sensor head temporarily enters a non-measurement state, for example, due to
insufficient received light amount or a RESET input state.
Example: When skipping occurs in the measurement values due to low incident level
When performing hold measurement, the output before the initial hold value is applied is the same as that for “Clamp”
even if “Keep” is set.
Multi View Explore : [Bank Group] | [(Bank Data Name)] (double click)
Edit pane : [I/O Settings] icon ( )
I/O Setting Screen : [Non-Measurement Setting]
Item Setting item Setting value Description
Non-Measure-
ment setting
Non-
Measurement
setting
Keep The measured value before the non-measurement state
w
as entered is held and output.
Clamp (default value) With analog output, the preset clamp value (abnormal
value)
is output. All judgment outputs become OFF.
1
Set the operating mode to the FUNC mode.
Displacement Sensor ZW-7000 series Confocal Fiber Type Displacement Sensor User's Man-
ual (Z362) "3-3 Switching operation modes"
(1)
(1)
(2)
(2)
(3)
(3)
Direction of
workpiece
movement
Amount of displacement
Amount of displacement
Skipping has occurred due to low
incident level
As the value before skipping occurs is kept, this is
effective in canceling measurement value skipping.
When [Keep] is selected