Recovering calibration ROM data
174
ZW-7000/5000
User's Manual
5-3 Recovering calibration ROM data
If an abnormality occurs in the sensor's calibration ROM, you can recover the backed up calibration ROM data
into the sensor.
Multi view Explore : [Device Group] | [(Sensor Name)] (double click)
Edit pane : [Tool] icon ( )
The calibration ROM data is different for each sensor serial number. Select the backup file that matches the sensor
serial number. Measurement will not be correct unless they match.
As a temporary measure if the Calibration ROM is lost, or fails, the measureme
nt can be restarted using the last
loaded Calibration ROM data according to the following procedure.
<Operating procedures>
Hold the button while [NO.ROM] is displayed on the sub digital. Press the ZERO/SET key after [OK/CAN] is
displayed.
<Note>
Confirm that the serial number of the Calibration ROM which is loaded is the same as the serial number of the
Sensor Head by checking the Controller Information.
If the serial numbers are not the same, measurement cannot be perf
ormed correctly.
When restarting the Sensor Controller, perform the same procedures again.
For a Displacement Sensor which has never been inserted or launche
d, this operation is disabled.
1
Set the operating mode to the FUNC mode.
3-2 Switching operation modes p.90
2
Click the [ROM recovery] in [Sensor setup].
Select the calibration ROM backup data file.