Wiring
72
ZW-7000/5000 
User's Manual
Analog output terminal block
Used for analog output.
Compatible cable specifications: Solid wire 0.2 to 1.5 mm
2 
Stranded wire 0.2 to 1.5 mm
2 
Stranded wire with rod terminal(no plastic sleeve) 0.25 to 1.5 mm
2 
Stranded wire with rod terminal(with plastic sleeve) 0.25 to 0.75 mm
2 
AWG 24 to 16 
Pin processed length 10 (±0.5) mm
• Cut the unnecessary signal lines so that they do not touch other signal lines.  
Make the wire length as short as possible. Do not use a wire length of ov
er 30 m.
• Analog output is not possible under the followin
g combination of conditions.
Measurement cycle is less than 40 µs.
EtherCAT output is enabled.
• Make sure to fix the terminal block to prevent be injure when pushing the release button with a screw driver.
Class Name Description
Analog  
output
OUT (V) This outputs the measured value, from -10 V to +10 V as the voltage value.
When measurement not possible: Approx. +10.8 V (default 
value; can be selected by user)
Alarm: Approx. 10.8 V
OUT (A) This outputs the measured value, from 4 mA to 20 mA as the current value.
When measurement not possible: Approx. +20.8mA (default value; can be selected by 
user)
Alar
m: Approx. +20.8 mA
OUT 0V 0 V for analog voltage output.
Signal name No. Signal name
OUT(V) 1 4 NC
OUT(A) 2 5 NC
OUT 0V 3 6 NC
Important