Basic Operations for Displaying Measurement Results
30
ZW-7000/5000
User's Manual
24 V input terminal block
Used for 24 VDC power supply.
Compatible cable specifications: Solid wire 0.2 to 2.5 mm
2
Stranded wire 0.2 to 2.5 mm
2
Stranded wire with rod terminal(no plastic sleeve) 0.25-2.5 mm
2
Stranded wire with rod terminal(with plastic sleeve) 0.25-2.5 mm
2
AWG 24-12
Pin processed length 10(±0.5) mm
1
Remove 24 V input terminal block from the Sensor
Controller.
2
Insert the wire in to the terminal hole.
•When you use a wire attached rod terminal,
you can insert just with Push-IN.
No. Signal name
1 DC24V (- )
2 DC24V (+)
Class Name Description
Power Supply DC24V (+) Terminal for external power supply (24 V)
DC24V (- ) Terminal for external power supply (0 V)