Cleaning the fiber connector and fiber cable using ZW-XCL
For ZW-7000 Series
The cleaner can only be used one time. Reusing the cleaner may result in dirt or scratches on the end surface,
causing malfunctions or measurement errors. Discard after use.
When using a fiber cleaner, be sure to use the cloth (white end surface
) on the cleaning section to clean.
1
Taking out a cleaner
Don't dirty the top of the cleaner, and take one cleaner
from t
he package.
2
Cleaning the fiber connector face of the Controller side.
Open the shtter of the fiber connector, and then wipe the fiber's face around thres reciprocating.
3
Cleaning the Sensor Head connection of the fiber ca-
ble
Remove the cap from the fiber cable and wipe the entire
e
nd surface of
the fiber connector with about 3 passes.
Clean around the 2 guide pins as well.