Setting the Output Conditions
ZW-7000/5000
User's Manual
131
4
Settings for Function
Multi View Explore : [Bank] | [(Bank Data Name)] (double click)
Edit pane : [Task Settings] icon ( )
Task Settings window : [Scaling]
Automatically Setting Scaling
Performing 1-point Scaling
Measurement is performed at one position, and offset values are set for that measured value.
The offset and increment/decrement inversion can be set.
Example: When displaying the distance to the sensing object
1 Select [Scaling].
2 Enter the correction value to [Span] and
[Offset].
Fixed scaling can also be executed by the operating keys on the Sensor Controller.
Performing Manual Scaling p.221
Item Setting item Setting value Description
Actual scaling
(one point)
Set value -999.999999 to 999.999999 [mm] Correct the measured value, and set the value to
display.
Scaling direction Inverse change (Far side is positive) The FAR side becomes a plus value.
Forward change (Near side is positive) The NEAR side becomes a plus value.
Note
4.5
4.5
Sensor head
Measured value:
4.5 mm
0 mm
Measuring center
distance
Sensing
object
Target display value:
Relative reference value
from the sensor with this
distance as the reference.
Default
display
Display value
(mm)
Measured value
(mm)