Setting the Output Conditions
132
ZW-7000/5000
User's Manual
Multi View Explore : [Bank] | [(Bank Data Name)] (double click)
→ Edit pane : [Task Settings] icon ( )
→ Task Settings window : [Scaling]
1 Set the [Scaling] to OFF.
2 Click [Auto setting].
The [Scaling] popup menu appears.
3 Select [Actual scaling (one point)], and click
[Next >].
The [Scaling] popup menu display changes to
[
1st poi
nt settings], and the current measured
value is displayed at [Current value].
4 Click [STOP] to fix the [Current value], set
[Set value] and click [Next >].
The [Scaling] popup menu display changes to
[
Direct
ion settings].
5 Set the change direction, and click [Next >].
The [Scaling] popup menu display changes to
[Conf
irm the scaling parameters], and the span
and offset values are displayed.
6 Click [OK].
1-point scaling can also be executed by the operating keys on the Sensor Controller.
Performing 1-point Scaling p.222
The [Current value] display is refreshed at a 100
ms cycle. When [STOP] is clicked, refreshing of
the [Current value] display stops. Clicking the
[STOP] button again resumes display refreshing.
Note