Setting the Output Conditions
ZW-7000/5000
User's Manual
133
4
Settings for Function
Performing 2-point Scaling
Measurement is performed at two positions, and offset values are set for those measured values.
Example: When correcting display values to match actual distances
Separate the two specified points by at least 1% of the rated measuring range for the connected Sensor Head.
For example, for the ZW-S7010, the two measured points must be sepa
rated by at least 1 mm × 0.01 = 0.01 mm as
the measuring range is 1 mm (± 0.5 mm).
Multi View Explore : [Bank] | [(Bank Data Name)] (double click)
→ Edit pane : [Task Settings] icon ( )
→ Task Settings window : [Scaling]
Item Setting item Setting value Description
Actual scaling
(two points)
Set value
(1st point)
-999.999999 to 999.999999 [mm] Correct the measured value of the 1st point, and set the
value to display.
Set value
(2nd point)
-999.999999 to 999.999999 [mm] Correct the measured value of the 2nd point, and set the
value t
o display.
1
Set the [Scaling] to OFF.
2
Click [Auto setting].
The [Scaling] popup menu appears.
4.0
-3.0
-2.8
3.8
-3.0
4.0
Display value
(mm)
Measured value
(mm)
Point 2
Point 1
Important