Setting the Output Conditions
134
ZW-7000/5000
User's Manual
3
Select [Actual scaling (two points)], and
click [Next >].
The [Scaling] popup menu display changes to
[
1st poi
nt settings].
4
Set correction of the 1st point.
Click [STOP] to fix the [Current value], set
[Set valu
e] and click [Next >].
The [Scaling] popup menu display changes to
[
2nd p
oint settings].
5
In the same way, set correction of the 2nd
point.
When the change direction is set and [Next>] is
c
licked,
the [Scaling] popup menu display
changes to [Confirm the scaling parameters],
and the span and offset values are displayed.
6
Click [OK].
2-point scaling can also be executed by the operating keys on the Sensor Controller.
Performing 2-point Scaling p.223
The current measured value is displayed at
[Current value], and is refreshed at 100 ms
cycles. When [STOP] is clicked, refreshing of the
[Current value] display stops. Clicking the
[STOP] button again resumes display refreshing.
Note
Note