Setting the Output Conditions
222
ZW-7000/5000
User's Manual
Automatically Setting Scaling
Performing 1-point Scaling
Performing 1-point Scaling p.131
Operating procedure
Steps Key operation Display Description
1 to 5 For moving to “SCALE”, see steps 1 to 5 in p.221.
6 Select “AUTO” as the scaling type.
The current setting value is displayed on the sub-display.
Press / key to enter editing mode and the sub-display blinks.
Select “AUTO” and press the
key.
7 Select “1POINT”.
Press the / keys to select “1PT” and press the key.
8 Press the key when the main display shows “POINT1”.
9 The current measured value is displayed on the main display and the
current setting value is displayed on the sub-display.
To edit numerical values, use the / / / keys.
Enter the setting value and then press the key.
10 Set the increment/decrement direction.
“DIRECT” is displayed on the main display and the incremental/
decremental
directions (“FWD”: the NEAR side as + and “REV”: the
FAR side as +) are displayed on the sub-display.
Press / key to enter editing mode and the sub-display blinks.
Enter the setting value and then press the key.
11 “OK/CAN” is displayed on the sub-display.
Press the key to execute scaling or key to cancel.
12 Press the key for two seconds to enter the RUN mode.