Basic Operations for Displaying Measurement Results
28
ZW-7000/5000
User's Manual
l
ZW-5000 Series
When connecting the fiber cable, mate the convex section of the fiber connector and the
groove on the sensor controller fiber adapter.
Mate the convex section of the fiber connector with
the g
roove on the fiber adapter and turn
the threaded section clockwise while pushing in (see figure 1).
In the "No good" status in figure 2 and figure 3, optic signa
ls cannot be transmitted and correct measure-
ment is not possible. Always check that the system is in the "Good" status.
Before connecting or disconnecting the calibration ROM, make sure that the Sensor Controller's power supply is
turned OFF. The Sensor Controller may break down if the calibration ROM is connected while the power is ON.
Use with the Calibration ROM always connected. If the C
alibration ROM is not connected, the error message
“SYSERR” is shown in the main digital and “NO.ROM” is shown in the sub digital, where measurement is not possi-
ble.
If the Calibration ROM is lost, or fails, refer to p.174.
6-1 Error Messages Errors Common to All Communication States in Displacement Sensor ZW-7000/
5000 series Confocal Fiber Type Displacement Sensor User’s Manual for Communications Settings
(Z363).
Be sure to use a Sensor Head and Calibration ROM with the same serial number.
A pair with different serial numbers cannot operate normally.
Make sure to initialize the Sensor Controller when connecting a different type
Sensor Head from that which was
connected the previous time the Sensor Controller was launched.
3
Connect the calibration ROM to the Sensor Control-
ler, and then tighten the supplied M2 screw.
Tightening torque: 0.15 N · m max.
Fiber connector
Fiber adapter
Groove
section
Convex
section
Threaded
section
(figure 2)
(figure 3)
figure 1
M2 screw
Calibration ROM
ROM connector
Important