R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 1065
tables at the beginning of the [SENSe:]FREQuency section.
In contrast to [SENSe<Ch>:]FREQuency:STARt? STOP. etc., SYSTem:FREQuency? can be used
in all sweep modes.
Center frequency of the sweep.
Device-specific, query only
See [SENSe<Ch>:]FREQuency:CENTer
SYSTem:IDENtify:FACTory
Resets the response to the *IDN? query to the factory default value. This command overwrites a user-
defined identification string; see example.
SYSTem:IDENtify:STRing 'MyDevice'; *IDN?
Define an identity string. The response is 'MyDevice'.
SYSTem:IDENtify:FACTory; *IDN?
Re-activate the factory setting. The response is of the form
'Rohde&Schwarz,ZVA8-4Port,12345,0.10.1.23', where ZVxx-nPort is the
analyzer type 12345 is the serial number of the analyzer and 0.10.1.23 is the
firmware version number.
SYSTem:IDENtify:STRing '<string>'
Defines an identification string (remote ID String) for the network analyzer. The query is equivalent to
*IDN?.
String variable for the name and directory of a NWA setup file (*.msi).
– (the function is not affected by *RST)
Device-specific, command or query
See SYSTem:IDENtify:FACTory
SYSTem:KLOCk <Boolean>
Locks or unlocks the local controls of the analyzer. This includes the front panel keys, the keyboard, or
other local interfaces.
ON | OFF – Lock or unlock the local keys.