R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 647
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DELTa:STATe <Boolean>
[Deprecated command]
Switches the delta mode for marker <Mk> on trace no. <Chn> on or off.
Note: This command is the ZVR-compatible equivalent of
CALCulate<Chn>:MARKer<Mk>:DELTa[:STATe].
Channel number used to identify the active trace
Marker number in the range 1 to 10
ON | OFF - Enables or disables the delta mode
Device-specific, command or query.
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <range_no>
Assigns a search range no. <numeric_value> to marker no <Mk> and selects the search range, e.g. in
order to display range limit lines or define the start and stop values.
Channel number used to identify the active trace
Marker number in the range 1 to 10
Number of the search range
0 – fixed full span search range (equal to the sweep range)
1 to 10 – user-definable search ranges; see example
0 (reserved for full span search range)
Device-specific, command or query.
CALC1:MARK1:FUNC:DOM:USER 2
Select the search range no. 2, assigned to marker no. 1 and trace no. 1.
CALC:MARK:FUNC:DOM:USER:STARt 1GHz
Set the start frequency of the search range to 1 GHz.
CALC:MARK:FUNC:DOM:USER:STOP 1.2GHz
Set the stop frequency of the search range to 1.2 GHz.
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER:SHOW <Boolean>
Displays or hides range limit lines for the search range selected via
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <range_no>.
This command is not available if the search range is Full Span (<range_no> = 0).
Channel number used to identify the active trace
Marker number in the range 1 to 10
ON | OFF - Range limit lines on or off