R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 648
Device-specific, command or query.
See CALCulate<Chn>:STATistics:DOMain:USER
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER:STARt <start_value>
Defines the start value of the search range selected via
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value>.
Channel number used to identify the active trace
Marker number in the range 1 to 10
Beginning of the search range.
Maximum allowed sweep range, depending on the instrument model and on the
sweep type. [Hz, dBm or s, depending on the sweep type]
Device-specific, command or query.
See CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value>
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER:STOP <stop_value>
Defines the stop value of the search range selected via
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value>.
Channel number used to identify the active trace
Marker number in the range 1 to 10
Maximum allowed sweep range, depending on the instrument model and on the
sweep type. [Hz, dBm or s, depending on the sweep type]
Device-specific, command or query.
See CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value>
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute MAXimum | MINimum | RPEak
| LPEak | NPEak | TARGet | LTARget | RTARget | BFILter | SPRogress
Selects a search mode for marker no. <Mk> and initiates the search. The marker must be created before
using CALCulate<Chn>:MARKer<Mk>[:STATe] ON (exception: bandfilter search).
Channel number used to identify the active trace
Marker number in the range 1 to 10. For a bandfilter search (BFILter) this
numeric suffix is ignored and may be set to any value because the bandfilter
search functions always use markers M 1 to M 4.