R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 732
DIAG:ALC:POFFset ON
Change last setting; cause the analyzer to use the ALC offset of the previous
measurement.
DIAGnostic:ALC:BW <bandwidth>
[Deprecated command, replaced by SOURce<Ch>:POWer<Pt>:ALC:BANDwidth]
Selects the bandwidth in the ALC control loop. The setting takes effect when automatic bandwidth setting
is disabled.
20 Hz to 7.12 MHz [Hz]. For a list of possible values refer to ALC Config.
Device-specific, command or query.
See DIAGnostic:ALC:SETTings:AUBW
DIAGnostic:ALC:CLAMp <Boolean>
[Deprecated command, replaced by SOURce<Ch>:POWer<Pt>:ALC:CLAMp]
Suspends the ALC mechanism while the analyzer acquires measurement data (Clamp ALC during
Measurement).
ON | OFF - Clamp ALC... on or off.
Device-specific, command or query.
See DIAGnostic:ALC:SETTings:AUBW
DIAGnostic:ALC:PIParameter AUTO | MANual
[Deprecated command, replaced by SOURce<Ch>:POWer<Pt>:ALC:PIParameter]
Enables automatic or manual setting of the PI controller parameters.
AUTO – Automatic setting, depending on the bandwidth (DIAGnostic:ALC:BW)
MANual – Manual setting using DIAGnostic:ALC:PIParameter:GAIN and
DIAGnostic:ALC:PIParameter:ITIMe.
Device-specific, command or query.
See DIAGnostic:ALC:SETTings:AUBW