R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 822
Parameters in
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:<std_type>
Sliding match (m) or
sliding match (f)
'<Ckit_Name>' ... <Max_Freq>[, <Port_1>]
no offset parameters, no load parameters (polynomial coefficients), no
OPEN | SHORt| MATCh
Reflect (m) or reflect (f)
'<Ckit_Name>' ... OPEN | SHORt[, <Port_1>]
no loss, otherwise complete parameter list
MMTHrough | MFTHrough |
FFTHrough
Through (m - m) or
through (m - f) or through
(f - f)
'<Ckit_Name>' ... <Loss>[, <Port_1>, <Port_2>]
no load parameters (polynomial coefficients), no OPEN | SHORt|
MATCh
MMLine | MMLINE1 | MFLine
| MFLINE1 | FFLine |
FFLINE1
MMLine(P2P3) ...
Line1 (m - m) or line1 (m
- f) or line1 (f - f)
Line1 (m - m, with port
restriction)...
'<Ckit_Name>' ... <Loss>[, <Port_1>, <Port_2>]
no load parameters (polynomial coefficients), no OPEN | SHORt|
MATCh
MMLINE<No.> |
MFLINE<No.> |
FFLINE<No.>
Line<No.> (m - m) or
Line<No.> (m - f) or
Line<No.> (f - f), with an
index <No.> between 1
and 99.
Especially for TRL and
NIST Multiline TRL
calibration.
'<Ckit_Name>' ... <Loss>[, <Port_1>, <Port_2>]
no load parameters (polynomial coefficients), no OPEN | SHORt|
MATCh
MMATten | MFATten |
FFATten
Attenuation (m - m) or
attenuation (m - f) or
attenuation (f - f)
'<Ckit_Name>' ... <Max_Freq>[, <Port_1>, <Port_2>]
no offset parameters, no load parameters (polynomial coefficients), no
OPEN | SHORt| MATCh
MMSNetwork | MFSNetwork |
FFSNetwork
Symmetric network (m -
m) or symmetric network
(m - f) or symmetric
network (f - f)
'<Ckit_Name>' ... OPEN | SHORt| MATCh[, <Port_1>, <Port_2>]
no loss, otherwise complete parameter list
[SENSe<Ch>:]CORRection:CKIT:DELete '<ckit_name>'
Deletes an imported or user-defined cal kit which is identified by its name. It is not possible to delete
predefined or ideal kits.
The result is undefined if multiple calibration kits with the same name but different labels exist.
Channel number. This suffix is ignored because calibration kits are channel-
independent.
String parameter containing an imported or user-defined calibration kit
available on the analyzer.
– (A *RST does not affect calibration kits)
Device-specific, no query
CORR:CKIT:N50:FOPEN 'NewKit1','',0,1.8E+010,0.0151,0,0,0.22,-0.22,0.0022
Create a new cal kit NewKit1 and assign an open (f) standard for the N 50 Ω
connector type with specific properties.