CORR:CKIT:DEL 'NewKit1'
Delete the previously created kit.
[SENSe<Ch>:]CORRection:CKIT:INSTall '<file_name>'
Loads cal kit data from a specified ZVR cal kit file.
Channel number. This suffix is ignored because calibration kits are channel-
independent.
String parameter to specify the name and directory of the cal kit file to be loaded.
Note: The loaded file must be a ZVR-specific cal kit file with the extension *.ck.
NWA cal kit files (*.calkit) can be imported using the MMEMory:LOAD:CKIT
command. Agilent cal kit files can be imported manually and converted into *.calkit
files.
Device-specific, no query.
CORR:CKIT:INST 'C:\Program Files\Rohde&Schwarz\Network
Analyzer\Calibration\Kits\ZCAN.ck'
Load the previously created ZVR cal kit file ZCAN.ck from the
default cal kit directory.
MMEM:STOR:CKIT 'ZCAN', 'C:\Program
Files\Rohde&Schwarz\Network
Analyzer\Calibration\Kits\ZCAN.calkit'
Store the imported cal kit data to a NWA cal kit file
ZCAN.calkit (assuming that the cal kit name stored in ZCAN.ck
reads ZCAN).
[SENSe<Ch>:]CORRection:CKIT:LABel '<ckit_name>', '<label>'
[SENSe<Ch>:]CORRection:CKIT:LABel? '<ckit_name>'
Assigns a label to an imported or user-defined calibration kit.
The result is undefined if multiple calibration kits with the same name but different labels exist.
Channel number. This suffix is ignored because calibration kits are channel-
independent.
String parameter containing an imported or user-defined calibration kit
available on the analyzer.
String parameter containing the label.
– (A *RST does not affect calibration kits)
Device-specific, command or query