R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 850
ON | OFF - Activates or deactivates calibrations (system error correction) at all
selected intermodulation frequency ranges.
Device-specific, command or query
CORR:COLL:IMOD ON
Activate system error correction at all selected intermodulation frequency
ranges
CORR:COLL:AUTO:PORT:TYPE FNPort, '', 1, 1, 2, 2
Select and initiate a full two port automatic calibration at analyzer ports 1 and 2
using ports 1 and 2 of the calibration unit
CORR:COLL:IMOD OFF
Deactivate system error correction at all selected intermodulation frequency
ranges
[SENSe<Ch>:]CORRection:COLLect:METHod TOM | TRM | TRL | TNA | TOSM |
FOPORT1 | FOPORT2 | FOPORT12 | FOPTport | ROPTport | REFL1 | REFL2 |
REFL12 | FRTRans | FTRans | RTRans | TPORt | UOSM
[Deprecated command]
Selects a one-port or two-port calibration type for channel <Ch> at ports 1/2.
Use the generalized command [SENSe<Ch>:]CORRection:COLLect:METHod:DEFine to select
the calibration type for arbitrary analyzer ports or a multiport calibration type.
Calibration types, see list below.
The numbers in the parameters denote the analyzer ports. Parameters for two-port
calibration types contain no numbers because the command is only valid for ports 1
and 2.
UOSM is interpreted as a mixer vector calibration if vector mixer measurements are
active ([SENSe<Ch>:]FREQuency:CONVersion VMIXer).
Confirmed (with device-specific calibration types), command or query. If several
"merged" calibration types are assigned to channel <Ch> (see
MMEMory:LOAD:CORRection:MERGe), the query returns a list of all calibration
types.
See [SENSe<Ch>:]CORRection:COLLect:SAVE.
The supported calibration types are listed below.
Type (Add Calibration dialog)