1 – Auto Length (and Loss), depending on the last
[SENSe<Ch>:]CORRection:COLLect:FIXTure:LMParameter:LOSS[:STATe]
setting
0 – Direct Compensation
Device-specific, command or query
See [SENSe<Ch>:]CORRection:COLLect:FIXTure[:ACQuire]
[SENSe<Ch>:]CORRection:COLLect:FIXTure:SAVE
Completes a fixture compensation, storing and applying the acquired data.
Device-specific, no query
See [SENSe<Ch>:]CORRection:COLLect:FIXTure[:ACQuire]
[SENSe<Ch>:]CORRection:COLLect:FIXTure:STARt
Prepares the analyzer for fixture compensation comprising a single or a series of fixture compensation
sweeps ([SENSe<Ch>:]CORRection:COLLect:FIXTure[:ACQuire]). Previous compensation data
is deleted.
Device-specific, no query
See [SENSe<Ch>:]CORRection:COLLect:FIXTure[:ACQuire]
[SENSe<Ch>:]CORRection:COLLect:IMODulation[:STATe] <Boolean>
Activates or deactivates calibrations (system error correction) at all selected intermodulation frequency
ranges.
This command has to be used with parameter ON when collecting calibration data for system error
correction for an intermodulation measurement. It ensures that the calibration will be performed at the
proper frequencies. It must be called before the first command of the calibration procedure (e.g. before
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine or
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure).
Please note that all calibrations performed after this command has been called with parameter ON will be
performed at all selected intermodulation frequency ranges. To prevent unwanted effects in non
intermodulation calibrations, the state should be reset to OFF after use.