Operating Manual 1145.1084.12 – 30 874
source port, assuming that the transmitted source power a
1
is correct. No
external cabling is needed.
CORR:POW:DATA? 'AWAV1'
Query the correction values. The analyzer returns 10 comma-separated
real numbers.
CORR:POW:DATA 'AWAV1', 1, 2, 3, 4, 5, -6, -7, -8, -9, -0
Replace the correction values by ten (new) numbers.
[SENSe<Ch>:]CORRection:POWer<port_no>:HARMonic:ACQuire
Starts the receiver calibration (3rd power calibration step for harmonic measurements), stores and applies
the calibration data.
The receiver calibration relies on the source power calibration acquired in steps no. 1 and 2.
[SENSe<Ch>:]CORRection:POWer<port_no>:HARMonic:ACQuire is enabled only after the
previous steps have been carried out
(SOURce<Ch>:POWer<port_no>:CORRection:HARMonic[:ACQuire]).
Calibrated channel number
Port number. This suffix is ignored; the analyzer calibrates the port that is selected as
a receive port for the harmonic measurement
([SENSe<Ch>:]FREQuency:CONVersion:HARMonic:RPORt).
Device-specific, no query
See SOURce<Ch>:POWer<port_no>:CORRection:HARMonic[:ACQuire].
[SENSe<Ch>:] CORRection:POWer<port_no>:IMODulation:ACQuire
[Deprecated command]
Starts the receiver calibration (3rd legacy power calibration step for intermodulation measurements),
stores and applies the calibration data.
The receiver calibration relies on the source power calibration acquired in step no. 1 using
SOURce<Ch>:POWer<port_no>:CORRection:IMODulation:LTONe[:ACQuire]. Make sure to first
perform step no. 1 before performing this step.
This command will perform a calibration only at the lower tone frequency. If the intermodulation
measurement is combined with a mixer measurement the calibration will be performed at the IF range of
the lower tone frequency only.
To perform a receiver calibration at the upper tone frequency and at all selected intermodulation
orders as well, use the
command[SENSe<Ch>:]CORRection:POWer<port_no>:IMODulation:RPORt:ACQuire
To perform a source power calibration of the receive port of the intermodulation measurement
(usually done as preparation for an enhanced wave correction) at the lower and upper tone
frequency and at all selected intermodulation orders, use the command