Improving level performance
R&S
®
SMB100B
162User Manual 1178.3711.02 ─ 08
The editor for user correction data provides a table with RF frequency and power
values and standard navigation functions.
The remote commands required to define the user correction data are described in
Chapter 12.15.2, "SOURce:CORRection subsystem", on page 446.
All columns in a row must contain values. Cells with missing values are therefore filled
automatically, using the value of the previous row.
If you use global dwell time in list mode, consider also that the instrument uses the
value set with Global Dwell Time for all list steps and not the values from the list.
Since the table and navigation functions can be assumed to be known, the following
description contains a brief overview, shown by the example of the "Edit List Mdoe
Data" dialog. If a function relates to a particular dialog, it is explicitly stated.
Settings
Edit List Mode Data.....................................................................................................162
Data handling keys......................................................................................................163
Go To............................................................................................................ 163
Edit................................................................................................................163
Fill with Sensor..............................................................................................163
Save As/Save............................................................................................... 163
Fill................................................................................................................................163
Edit List Mode Data
Table with values for list or user correction processing.
Note: Once you enter a value, the editor automatically adds preset values in the other
columns. This functionality protects against data loss, otherwise incomplete rows are
lost when saving. You can simply override these values.
"Frequency /Hz"
Sets the frequency values.
Remote command:
[:SOURce<hw>]:LIST:FREQuency on page 479
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency
on page 449
"Power /dBm" (list mode data table)
Sets the level values.
Remote command:
[:SOURce<hw>]:LIST:POWer on page 480
"Correction Value /dBm" (user correction data table)
Sets the level values.
Remote command:
[:SOURce<hw>]:CORRection:CSET:DATA:POWer on page 449