Instrument Function
R&S
®
SMBV100A
434Operating Manual 1176.8016.02 ─ 17
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer:STEP
on page 628
Delay Start
Sets the delay of the carrier marked by "Carrier Start".
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:DELay[:STARt]
on page 627
Delay Step
Sets the step width that is used to increment the delay
The resulting delay in the carrier table equals:
DelayStart + n* DelayStep
where n ranges from 0 to (Carrier Stop – Carrier Start).
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:DELay:STEP
on page 626
Input Waveform File
Calls the "File" menu for selecting the inputfile with the I/Q signal to be modulated onto
all carriers of the selected carrier range.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:FILE on page 627
Apply Assistant Settings
Transfer the assistant settings to the carrier table.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:EXECute
on page 627
Carrier Table
The table displays the settings of all available carriers. Previously applied assistant set-
tings can be further refined. The number of lines corresponds to the number of carriers.
Note: The phase/deg settings are only valid if optimization of the crest factor is dis-
abled ("Crest Factor Mode" = Off).
No ← Carrier Table
Indicates the carrier index ranging from 0 to (number of carriers -1).
Individual carriers can be set using the remote control commands by specifying the
index in the parameter CARR.
Baseband Signal - Baseband Block